Abstract
A positive bronchodilator response has been defined as a 12% increase in the forced expiratory volume in one second (FEV
) or forced vital capacity (FVC) from their respective pre-bronchodilator values, combined with at least a 0.2 L absolute change. Recent recommendations suggested the use of the percent change in FEV
and FVC relative to their predicted normal values without having applied them in patients with airflow obstruction. The aim of the current study was to compare the two approaches over a wide range of pre-bronchodilator FEV
and FVC values.
A retrospective review of consecutive patients undergoing spirometry and bronchodilator testing was completed. The change in FEV
and FVC with a bronchodilator was expressed relative to the pre-bronchodilator and predicted normal FEV
and FVC.
In 1,040 patients with a non-paradoxical change in FEV
, 19.0% had a ≥ 12% change in FEV
using their pre-bronchodilator value compared to 5.7% using their predicted normal value. For FVC, the respective values were 12.7% vs. 5.8%. The difference was retained in patients with a ≥ 0.2 L change in FEV
or FVC. In unobstructed patients, the upper threshold (two standard deviations above the mean) of the bronchodilator response was 14% for FEV
and 10% for FVC using predicted normal values.
Expressing the percent change in FEV
and FVC relative to predicted normal values reduces the over-estimation of the bronchodilator response, especially in patients with a very low pre-bronchodilator FEV1, including in those with a ≥ 0.2 L change in FEV
. Irrespective of pre-bronchodilator values, a ≥ 14% change in FEV
and ≥ 10% change in FVC relative to the predicted normal values could be considered a positive bronchodilator response.