Abstract
We describe an automated, kinetic nephelometric method for fibronectin on the Multistat III F/LS Centrifugal Analyzer (Instrumentation Laboratory Inc., Lexington, MA 02173). Antiserum is diluted with polyethylene glycol. Calibrators and samples are prediluted with potassium phosphate buffer (10 mmol/L, pH 7.0) containing 8.5 g of NaCl per litre. Intensity (I) is read at 5 and 180 s and the resulting δI plotted against concentration. A non-linear least squares curve fitting and interpolation of results is carried out automatically.
Three controls, with values of between 190–370 gave coefficients of variation between 4 and 7 percent, and the sensitivity of the method is to 25 mg/L.