Validation of daily 0.35 T diffusion-weighted MRI for MRI-guided glioblastoma radiotherapy
Journal article   Open access   Peer reviewed

Validation of daily 0.35 T diffusion-weighted MRI for MRI-guided glioblastoma radiotherapy

Natalia Lutsik, Siamak P Nejad-Davarani, Alessandro Valderrama, Janette Herr, Danilo Maziero, Kaylie Cullison, Gregory A Azzam, Gregory J Kubicek, Jessica Meshman, Macarena I de la Fuente, …
Medical physics (Lancaster)
2024-04-08
PMID: 38588475

Abstract

MRI‐Linac diffusion‐weighted MRI glioblastoma
MRI-Linac systems enable daily diffusion-weighed imaging (DWI) MRI scans for assessing glioblastoma tumor changes with radiotherapy treatment. Our study assessed the image quality of echoplanar imaging (EPI)-DWI scans compared with turbo spin echo (TSE)-DWI scans at 0.35 Tesla (T) and compared the apparent diffusion coefficient (ADC) values and distortion of EPI-DWI on 0.35 T MRI-Linac compared to high-field diagnostic MRI scanners. The calibrated National Institute of Standards and Technology (NIST)/Quantitative Imaging Biomarkers Alliance (QIBA) Diffusion Phantom was scanned on a 0.35 T MRI-Linac, and 1.5 T and 3 T MRI with EPI-DWI. Five patients were scanned on a 0.35 T MRI-Linac with a TSE-DWI sequence, and five other patients were scanned with EPI-DWI on a 0.35 T MRI-Linac and a 3 T MRI. The quality of images was compared between the TSE-DWI and EPI-DWI on the 0.35 T MRI-Linac assessing signal-to-noise ratios and presence of artifacts. EPI-DWI ADC values and distortion magnitude were measured and compared between 0.35 T MRI-Linac and high-field MRI for both phantom and patient studies. The average ADC differences between EPI-DWI acquired on the 0.35 T MRI-Linac, 1.5 T and 3 T MRI scanners and published references in the phantom study were 1.7%, 0.4% and 1.0%, respectively. Comparing the ADC values based on EPI-DWI in glioblastoma tumors, there was a 3.36% difference between 0.35 and 3 T measurements. Susceptibility-induced distortions in the EPI-DWI phantoms were 0.46 ± 1.51 mm for 0.35 MRI-Linac, 0.98 ± 0.51 mm for 1.5 T MRI and 1.14 ± 1.88 mm for 3 T MRI; for patients -0.47 ± 0.78 mm for 0.35 T and 1.73 ± 2.11 mm for 3 T MRIs. The mean deformable registration distortion for a phantom was 1.1 ± 0.22 mm, 3.5 ± 0.39 mm and 4.7 ± 0.37 mm for the 0.35 T MRI-Linac, 1.5 T MRI, and 3 T MRI scanners, respectively; for patients this distortion was -0.46 ± 0.57 mm for 0.35 T and 4.2 ± 0.41 mm for 3 T. EPI-DWI 0.35 T MRI-Linac images showed higher SNR and lack of artifacts compared with TSE-DWI, especially at higher b-values up to 1000 s/mm . EPI-DWI on a 0.35 T MRI-Linac showed superior image quality compared with TSE-DWI, minor and less distortions than high-field diagnostic scanners, and comparable ADC values in phantoms and glioblastoma tumors. EPI-DWI should be investigated on the 0.35 T MRI-Linac for prediction of early response in patients with glioblastoma.
pdf
Validation of daily 0.35 T diffusion-weighted MRI for MRI-guided glioblastoma radiotherapy9.27 MBDownloadView
Open Access CC BY-NC-ND V4.0
url
https://doi.org/10.1002/mp.17067View
Published (Version of record) Open

Metrics

InCites Highlights

These are selected metrics from InCites Benchmarking & Analytics tool, related to this output

Citation topics
1 Clinical & Life Sciences
1.113 Brain Imaging
1.113.453 Glioma Research
Web Of Science research areas
Radiology, Nuclear Medicine & Medical Imaging
ESI research areas
Clinical Medicine

Details

Logo image