Abstract
BACKGROUND: Portable monitoring is a convenient means for diagnosing sleep apnea. However, data on whether one night of monitoring is sufficiently precise for the diagnosis of sleep apnea are limited.
RESEARCH QUESTION: The current study sought to determine the variability and misclassification in disease severity over three consecutive nights in a large sample of patients referred for sleep apnea.
METHODS: A sample of 10,340 adults referred for sleep apnea testing was assessed. A self-applied type III monitor was used for three consecutive nights. The apnea-hypopnea index (AHI) was determined for each night, and a reference AHI was computed by using data from all 3 nights. Pairwise correlations and the proportion misclassified regarding disease severity were computed for each of the three AHI values against the reference AHI.
RESULTS: Strong correlations were observed between the AHI from each of the 3 nights (r = 0.87-0.89). However, substantial within-patient variability in the AHI and significant misclassification in sleep apnea severity were observed based on any 1 night of monitoring. Approximately 93% of the patients with a normal study on the first night and 87% of those with severe sleep apnea on the first night were correctly classified compared with the reference derived from all three nights. However, approximately 20% of the patients with mild and moderate sleep apnea on the first night were misdiagnosed either as not having sleep apnea or as having mild disease, respectively.
CONCLUSIONS: In patients with mild to moderate sleep apnea, one night of portable testing can lead to misclassification of disease severity given the substantial night-to-night variability in the AHI.